Abstract

PurposeThe assessment of retinal nerve fiber layer defects (RNFLDs) is a useful part of glaucoma care. Here, we obtained en-face images of retinal layers below the inner limiting membrane (ILM) with swept source-optical coherence tomography (SS-OCT), and measured RNFLD angle with new software.MethodsThis study included 105 eyes of 105 normal tension glaucoma (NTG) patients (age, 59.8 ± 13.2). Exclusion criteria were best-corrected visual acuity < 0.5, axial length > 28 mm, non-glaucoma ocular disease, and systemic disease affecting the visual field. We obtained 12 x 9 mm 3D volume scans centered on the macula with SS-OCT (DRI OCT-1, Topcon), and from these scans, created 3 averaged en-face images, each comprising 7 horizontal en-face images (total thickness: 18.2 μm). We labeled these averaged images, according to their depth below the ILM, as en-face images 1 (shallowest), 2 (middle) and 3 (deepest). In each image, a circle was drawn centered on the disc, with a radius of half the distance between the centers of the disc and macula. The investigator marked points where the edge of the RNFLD intersected this circle, and RNFLD angle (RNFLDA) was calculated with new software. Finally, we analyzed the association between RNFLDA, cpRNFLT, weighted RGC count (wrgc) and Humphrey field analyzer (HFA)-measured mean deviation (MD) and hemifield total deviation (TD), both overall and in each hemifield.ResultsEn-face image 2 had the highest interclass reproducibility for measuring RNFLDA (intra-rater intraclass correlation coefficient (ICC): 0.988, inter-rater ICC: 0.962). The correlation coefficients with RNFLDA were: HFA MD, -0.60; superior TD, -0.73; inferior TD, -0.69; overall cpRNFLT, -0.27; superior hemifield cpRNFLT, -0.39; and inferior hemifield cpRNFLT, -0.53 (all p<0.001).ConclusionsRNFLDA measured in SS-OCT images had high reproducibility and was correlated to glaucoma severity. Our new method may be a valuable future part of glaucoma care.

Highlights

  • Glaucoma affects over 70 million people worldwide, and is the second most common cause of blindness [1,2]

  • RNFLD angle (RNFLDA) measured in swept source-optical coherence tomography (SS-optical coherence tomography (OCT)) images had high reproducibility and was correlated to glaucoma severity

  • OCT, a technology introduced by Huang et al [15], is often used to measure circumpapillary retinal nerve fiber layer thickness, a parameter that has become established as a useful part of early glaucoma diagnosis [11,12,16]

Read more

Summary

Methods

This study included 105 eyes of 105 normal tension glaucoma (NTG) patients (age, 59.8 ± 13.2). We obtained 12 x 9 mm 3D volume scans centered on the macula with SS-OCT (DRI OCT-1, Topcon), and from these scans, created 3 averaged en-face images, each comprising 7 horizontal enface images (total thickness: 18.2 μm). We labeled these averaged images, according to their depth below the ILM, as en-face images 1 (shallowest), 2 (middle) and 3 (deepest). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript This does not alter our adherence to PLOS ONE policies on sharing data and materials

Results
Introduction
Material and methods
Discussion
Full Text