Retinal images are obtained by simultaneously illuminating and imaging the retina, which is achieved using a fundus camera. This device meets low light illumination of the fundus with high resolution and reflection free images. Although the current equipment presents a sophisticated solution, it is complex to align due to the high number of off-axis components. In this work, we substitute the complex illumination system by a ring of LEDs mounted coaxially to the imaging optical system, positioning it in the place of the holed mirror of the traditional optical design. We evaluated the impact of this substitution regarding to image quality (measured through the modulation transfer function) and illumination uniformity produced by this system on the retina. The results showed there is no change in image quality and no problem was detected concerning uniformity compared to the traditional equipment. Consequently, we avoided off-axis components, easing the alignment of the equipment without reducing both image quality and illumination uniformity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.