Abstract

The European Commission's Directorate General Joint Research Centre has organized several proficiency tests on the determination of acrylamide (AA) in food. This paper presents the results and outcome of a proficiency test that focused on the determination of AA in crispbread samples. One of the goals was the identification of the influence of different parameters such as analyte extraction or instrument calibration on the analytical results. A set of samples, containing 3 different crispbread samples as well as extracts of one crispbread sample and AA standard solutions, was shipped to each participant. A total of 42 European laboratories reported analytical results that were evaluated by applying internationally accepted protocols and procedures. The study found that, for each sample, the results of 4-8 laboratories were outside the range formed by the target value plus or minus the 2-fold of the target standard deviation; thus, they did not perform satisfactorily. In transferring this knowledge to the data of monitoring databases of AA in food, care must be taken that data are quality controlled, as it is likely that some of them may be biased.