Abstract

Pregnant women (n = 859) in rural Cameroonian prenatal clinics were screened by two rapid human immunodeficiency virus (HIV) antibody tests (rapid tests [RT]) (Determine and Hema-Strip) using either whole blood or plasma. One additional RT (Capillus, HIV-CHEK, or Sero-Card) was used to resolve discordant results. RT results were compared with HIV-1 enzyme immunoassay (EIA) and Western blot (WB) results of matched dried blood spots (DBS) to assess the accuracy of HIV RTs. DBS EIA/WB identified 83 HIV antibody-reactive, 763 HIV antibody-nonreactive, and 13 indeterminate specimens. RT results were evaluated in serial (two consecutive tests) or parallel (two simultaneous tests) testing algorithms. A serial algorithm using Determine and Hema-Strip yielded sensitivity and specificity results of 97.6% and 99.7%, respectively, whereas a parallel RT algorithm using Determine plus a second RT produced a sensitivity and specificity of 100% and 99.7%, respectively. HIV RTs provide excellent alternatives for identifying HIV infection, and their field performance could be monitored using DBS testing strategies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.