Abstract

Staff are exposed to radiation in the scanning room when assisting with CT scans of patients requiring ventilatory support during procedures. We measured lens doses using a phantom during a high-energy trauma protocol. Dosimetry showed that the unprotected lens received 2.02 mGy on the right and 1.91 mGy on the left, which are not negligible doses. Respective exposures to the right and left lens were 53.6% and 55.1% when wearing 0.07 mm Pb protective glasses with side covers; 53.7% and 64.2% when wearing 0.7 mm Pb glasses without side covers when facing away from the patient couch; and 92.1% and 91.2% using protective shielding in the gantry. Since the face direction may change during assistance with CT imaging, it is desirable that the protective glasses have a shape with a side cover. The protective shielding had a major radiation reduction effect, although it is expensive to acquire, install, and maintain.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.