Abstract

Staff are exposed to radiation in the scanning room when assisting with CT scans of patients requiring ventilatory support during procedures. We measured lens doses using a phantom during a high-energy trauma protocol. Dosimetry showed that the unprotected lens received 2.02 mGy on the right and 1.91 mGy on the left, which are not negligible doses. Respective exposures to the right and left lens were 53.6% and 55.1% when wearing 0.07 mm Pb protective glasses with side covers; 53.7% and 64.2% when wearing 0.7 mm Pb glasses without side covers when facing away from the patient couch; and 92.1% and 91.2% using protective shielding in the gantry. Since the face direction may change during assistance with CT imaging, it is desirable that the protective glasses have a shape with a side cover. The protective shielding had a major radiation reduction effect, although it is expensive to acquire, install, and maintain.