Abstract

In recent years, aortic aneurysm treatment with stent graft grafting in the X-ray fluoroscopy is increasing. This is an endovascular therapy, because it is a treatment which includes the risk of radiation damage, having to deal with radiation damage, to know in advance is important. In this study, in order to grasp the trend of exposure stent graft implantation in a hybrid operating room (OR) system, focusing on clinical data (entrance skin dose and fluoroscopy time), was to count the total. In TEVAR and EVAR, fluoroscopy time became 13.40 ± 7.27 minutes, 23.67 ± 11.76 minutes, ESD became 0.87 ± 0.41 mGy, 1.11 ± 0.57 mGy. (fluoroscopy time of EVAR was 2.0 times than TEVAR. DAP of EVAR was 1.2 times than TEVAR.) When using the device, adapted lesions and usage are different. This means that care changes in exposure-related factors. In this study, exposure trends of the stent graft implantation was able to grasp. It can be a helpful way to reduce/optimize the radiation dose in a hybrid OR system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.