Abstract

This prospective study was done to evaluate the accuracy of pulse oximetry as an indicator of hypoxemia in sickle cell anemia patients presenting to the ED in acute vasoocculsive crisis. Thirteen patients (age older than 18 years) presenting in crisis, with pulse oximetry readings less than 95% had arterial blood gas (ABG) analysis performed. The mean pulse oximetry reading was 91.4% ± 3.2% (SD); the mean ABG oxygen saturation (ABG O 2 sat) was 95.3% ± 1.4% (SD). The mean O 2 sat difference (ABG O 2 sat minus pulse oximetry) was 4.5% ( P < .001 versus a control group of ED patients receiving ABG and pulse oximetry). Pulse oximetry correlated with ABG O 2 sat ( r = 0.889, P < .001) and O 2 sat difference ( r = −0.927, P < .001). No correlation existed between complete blood count, reticulocyte count, historical and physical examination findings, and O 2 sat or O 2 sat difference. By linear regression, ABG O 2 sat = 0.44 (pulse oximetry) + 55.6. This study suggests that pulse oximetry underestimates oxygenation in sickle cell anemia patients presenting in painful crisis.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call