Abstract

To assess the clinical applicability of pulse oximetry in the intensive care setting, a comparison was made of arterial hemoglobin saturation values determined by in vitro oximetry (SaO2) and pulse oximetry (SpO2) in 21 critically ill dogs. Single SaO2 measurements were compared to simultaneously obtained SpO2 readings. The correlation between these two methods was statistically significant (r = 0.8944, p = 0.0001). In addition, heart rates read by the pulse oximeter were compared to simultaneously obtained electrocardiograms (ECG). The correlation between these two methods was statistically significant (r = 0.9966, p = 0.0001). The pulse oximeter was easy to use, and recorded trends in oxygenation virtually instantaneously. Pulse oximetry appears to be an accurate and practical technique for the continuous non‐invasive monitoring of oxygenation in critically ill dogs in the intensive care unit.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.