Abstract

A gamma ray detector for PET, consisting of an array of mixed lutetium oxyorthosilicate (MLS) scintillator crystals coupled to a position sensitive avalanche photodiode (PSAPD), was evaluated. The scintillator array was constructed from individual MLS crystals with dimensions of 1.5 mm/spl times/1.5 mm/spl times/15 mm. The assembled 7/spl times/7 array, including inter-crystal reflector material, had a pitch of 1.79 mm. The low noise, high gain PSAPD had dimensions of 14 mm/spl times/14 mm. Peaks associated with each of the 49 scintillator crystals were readily identifiable in flood histograms, and most of the crystals demonstrated energy resolution in the range of 15% to 20% at 511 keV. Preliminary measurements of the timing of the PSAPD in coincidence with a fast-scintillator/PMT detector indicated a timing resolution of approximately 4 ns. The operating characteristics and design attributes, such as compactness and reduced readout channel requirements, of the PSAPD make it attractive for high resolution PET applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.