Abstract
For evaluating the counting efficiency of a portal monitor, we use a137Cs radiation point source (1 μCi) to subsequently establish it effective measurable area. Through simulation, we estimate the appropriate distance from potentially contaminated individuals in the scanning queue to the monitoring individual. When this distance is over 10 m, the counting efficiency was below 0.01%. We find that the triage can be applied to roughly 180 individuals per hour during mass casualties.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.