Abstract
Patient-specific intensity-modulated radiation therapy (IMRT) quality assurance (QA) is essential for complex radiotherapy treatment as it involves complex intensity modulation and high-dose gradient regions. IMRT QA was performed by point dose verification and two-dimensional (2D) dose distribution measurement using gamma method. Calibrated External Beam Therapy 3 (EBT3) film was used for point dose and pre-treatment verification of 10 IMRT plans, five complex Head and Neck (HN) and five brain cases. The gamma passing rate (GPR) was evaluated for 3%/3mm gamma criteria and compared with 2D array. Isocentre dose was measured for all 10 IMRT plans on EBT3 film. Percentage deviation of point dose measurement from TPS calculated was found 0.4% for brain cases and 2.9% for HN cases. The GPR for 3%/3mm criteria was obtained higher than 95% for brain and HN cases. Results suggest that film dosimetry is also a reliable verification system for patient-specific IMRT QA as the 2D array.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.