Abstract
This study investigates patients’ dose levels during digital mammography procedures and proposes a local diagnostic reference level (DRL). The patient sample is 1055 patients who have been examined at the Riyadh Care Hospital, Riyadh, Saudi Arabia. All procedures were carried out for medically justified clinical conditions. All procedures were carried out using a direct digital mammography (DDR) radiology system. The mean and range of ESAK (mGy) were 5.19 ± 3.18 (0.33–29.9), while the mean AGD (mGy) is 1.3±(1.0–7.2) per procedure. The third quartile value for ESAK (mGy) and AGD (mGy) is 6 and 1.5, respectively. Measured ESAK (mGy) per procedure and estimated AGD are comparable with previous studies. The interpatient and interdepartmental variation is attributed mainly to the number of projections per procedure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.