Abstract

Summary A survey was performed to evaluate patient exposure after a procedure for the CT X-ray tube’s emergency warm-up, with the patient positioned on the table within the CT gantry. Three CT units, situated in three different University hospitals, were included in the survey. The evaluation was performed with AGFA personal monitoring films to visualize the results and discuss them with the radiographers who operate the systems. Additional measurements were performed with the RaySafe X2 system to demonstrate the presence of exposure. The air kerma resulting from implementing the warm-up protocol was evaluated to be higher than 112 μGy, 409 μGy, and 807 μGy for each of the CT units. Those values were underestimated because of the dosimetry equipment used and the methodology. A discussion between medical staff, engineers, and medical physicist was initiated. The practice with unnecessary patient exposure was terminated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.