Abstract

A new method is proposed for the spectrophotometric determination of Pd(II), based on the reaction of Pd(II) with 2-(4-chloro-2-phosphonophenylazo)-7-(3-carboxyphenylazo)-1,8-dihydroxynaphthalene-3,6-disulfonic acid(CPA-mK) in sulfuric acid without heating. Beer’s law is obeyed for 1.0–4.0 μg of Pd (II) in 10 mL of solution. The calibration curve from 1.0 to 42.0 μg in 10 mL of solution is modeled successfully by artificial neural networks (ANNs). The maximum relative error between experimental values and the values predicted by ANNs is 1.5%. In comparison with some mathematical functions, ANNs show better ability for curve fitting, thus greatly extending the applicable range of the calibration curve of this new system. The method has been applied to determine Pd (II) in ore and catalyst samples with a relative error of less than 4% and with a recovery range between 94% and 103%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.