Abstract
BackgroundReliable, accurate, noninvasive, and continuous determination of hemoglobin would be an important advance in the care of trauma patients. The aim of this study was to evaluate the utility of the Masimo Radical 7 device in severely injured trauma patients. MethodsHighest level trauma activation patients were enrolled over a 1-year period. Laboratory hemoglobin values were compared with Masimo hemoglobin values using Bland-Altman analysis. ResultsA total of 525 patients were enrolled in the study. Comparison of 861 paired values from 418 patients showed a variance of 3.89 to −3.84 g/dL, showing a nonsignificant correlation between Masimo hemoglobin and laboratory hemoglobin values. ConclusionsThe Masimo Radical 7 system evaluated in this study holds promise, but it is not ready to be used as an initial noninvasive evaluation tool in the acute treatment of severely injured trauma patients. There was a poor correlation between Masimo hemoglobin and laboratory hemoglobin and large numbers of missing data. On the basis of the poor correlation, the Masimo Radical 7 device cannot currently be used to guide transfusion therapy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.