Abstract

A novel enzyme-linked immunosorbent assay (ELISA [Dialbumin]) for rapid office measurement of microalbuminuria was evaluated and its performance compared with that of a commercially available radioimmunoassay (double-antibody albumin). Urine samples containing between 0.75 and 1800 micrograms/ml of albumin were obtained from 31 diabetic patients and assayed by both methods. A comparison of the paired values obtained from the two methods gave a correlation coefficient of greater than 0.99. The Dialbumin assay, which used detachable eight-well strips (1 strip/sample), 10-min incubation, tap water wash, and a 2-min color development step, was read on both an ELISA reader and a hand-held analytical device (Acc-U-Dial) designed specifically for this test. The findings of this study indicate that the Dialbumin assay, used in conjunction with the Acc-U-Dial device, affords a rapid, convenient, and sensitive method for quantitative determination of a broad range (0.3-1280 micrograms/ml) of urinary albumin levels in the office setting.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.