Abstract

Neutron production is of concern for proton therapy, especially for passive scattering proton beam delivery methods. The levels of neutron dose equivalent vary significantly with system design and treatment parameters. The purpose of this study was to examine neutron dose equivalent per therapeutic dose (H/D) around the Mevion S250 proton therapy system, a novel design of proton therapy systems. The benchmark comparisons between measurement and simulation were found to be within a factor of 2 for most cases. The H/D values were evaluated as a function of various parameters. The results showed that, at a standard reference condition (10 × 10 cm2 field size, distance 1 m detector-to-isocenter lateral to the primary proton beam direction), the H/D values range from 0.72 to 3.37 mSv Gy−1 for all configurations studied. The H/D values generally (1) decreased as the neutron detectors moved away from the isocenter, (2) decreased with increasing aperture field sizes, (3) increased with increasing angle from the initial beam axis and (4) were independent of treatment nozzle position. The H/D trends were consistent with other existing passive scattering proton accelerators reported in the literature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.