Abstract

To compare the effect of standardized upper extremity position versus varied upper extremity positions on neonatal peripherally inserted central catheter (PICC) tip placement and movement. Secondary outcomes assessed were compliance with education, implementation, and complication rates. Tip movement was analyzed between 136 post-PICC insertion x-ray pairs from 72 infants in the 6 months before and after standardization of upper extremity position. Tip movement was regressed over days between x-ray pairs, respiratory support, absolute weight change, and insertion vein. There was no difference in PICC tip movement among varied analysis pairs or when standard position pairs were compared with pairs that were in a same nonstandard position. Days between x-rays, respiratory support, absolute weight change, and insertion vein did not predict tip movement. Attendance at education sessions was 100%. Compliance with the new standard was 73%. Complication rates were not significantly different. Standardization of upper extremity position during neonatal PICC confirmation x-rays did not alter PICC tip movement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.