Abstract

1. Multi-frequency bio-impedance analysis has been used to estimate extracellular and total body water in a heterogeneous group of 43 surgical patients (23 males, 20 females). 2. Radioisotope-dilution methods were used for the measurement of extracellular and total body water. 3. Resistance and reactance were measured between wrist and ankle at frequencies from 5 kHz to 1 MHz. 4. Extracellular and total body water were estimated by multiple stepwise regression using the radioisotope values as the dependent variables. The parameters included in the regression were: resistance and reactance at each frequency, body habitus parameters, plasma albumin and plasma sodium. 5. The standard errors of the estimates between the measured and estimated values were 1.73 litres (coefficient of variation 9.6%) and 2.17 litres (coefficient of variation 6.0%) for extracellular and total body water, respectively. 6. These errors represent a useful improvement relative to those obtained from anthropometric estimates. However, the improvements relative to the use of a single frequency (50 kHz) are not clinically significant.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.