Abstract

Manual reading of fluorescent acid-fast bacilli (AFB) microscopy slides is time-intensive and technically demanding. The aim of this study was to evaluate the accuracy of MetaSystems' automated fluorescent AFB slide scanner and analyzer. Auramine O-stained slides corresponding to 133 culture-positive and 363 culture-negative respiratory (n = 284), tissue (n = 120), body fluid (n = 81), and other (n = 11) sources were evaluated with the MetaSystems Mycobacteria Scanner running the NEON Metafer AFB Module. The sensitivity and specificity of the MetaSystems platform was measured as a standalone diagnostic and as an assistant to technologists to review positive images. Culture results were used as the reference method. The MetaSystems platform failed to scan 57 (11.5%) slides. The MetaSystems platform used as a standalone had a sensitivity of 97.0% (129/133; 95% CI 92.5 to 99.2) and specificity of 12.7% (46/363; 95% CI 9.4 to 16.5). When positive scans were used to assist technologists, the MetaSystems platform had a sensitivity of 70.7% (94/133; 95% CI 62.2 to 78.3) and specificity of 89.0% (323/363; 95% CI 85.3 to 92.0). The manual microscopy method had a sensitivity of 79.7% (106/133; 95% CI 71.9 to 86.2) and specificity of 98.6% (358/363; 95% CI 96.8 to 99.6). The sensitivity of the MetaSystems platform was not impacted by smear grade or mycobacterial species. The majority (70.3%) of false positive smears had ≥2+ smear results with the MetaSystems platform. Further performance improvements are needed before the MetaSystems' automated fluorescent AFB slide reader can be used to assist microscopist in the clinical laboratory.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call