Abstract

Information regarding the use of MALDI-TOF MS as an alternative to conventional laboratory methods for the rapid and reliable identification of bacterial isolates is still limited. In this study, MALDI-TOF MS was evaluated on 295 anaerobic isolates previously identified by 16S rRNA gene sequencing and with biochemical tests (Rapid ID 32A system, BioMérieux). In total, 85.8% of the isolates were identified by MALDI-TOF MS at the species level vs 49.8% using the Rapid ID 32A system (p < 0.0001). None of the isolates was discordantly identified at the genus level using MALDI-TOF MS and only 9 of them could not be identified using the method. Thus, our results show that MALDI-TOF MS is a robust and reliable tool for the identification of anaerobic isolates in the microbiology laboratory. Its implementation will reduce the turnaround time for a final identification and the number of isolates that require 16S rRNA sequencing.