Abstract
Matrix-Assisted Laser Desorption Ionisation-Time of Flight (MALDI-ToF) Mass Spectrometry with Bruker MALDI Biotyper software was evaluated as a method for identifying veterinary bacteria. For 620 isolates (~100 bacterial species), identification by MALDI-ToF and non-16S rDNA methods (mainly phenotypic/biochemical) agreed to species-level (95.3%) and to species/genus-level (100%), but in the absence of 16S rDNA as a gold standard. For a further panel of 107 anaerobes and 234 aerobes (~100 bacteria species) using 16S rDNA results as the gold standard, MALDI-ToF/biochemical tests showed 97.8/96.6% species-level and 99.6/93.5% genus-level agreement for aerobes and 95.3/93.6% species-level and 100/95.3% genus-level agreement for anaerobes compared to the gold standard. Where results were obtained from direct spots, direct spots overlaid with formic acid and extracts, 89.4% of 180 aerobes and 90.1% of 152 anaerobes were identified by MALDI-ToF. MALDI-ToF was shown to be a rapid and reliable method to identify veterinary bacteria.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.