Abstract

Five hundred specimens (288 genital, 192 dermal, and 20 ocular) were extracted by technologists, and the DNA was assayed by LightCycler PCR (DNA polymerase and thymidine kinase [TK] gene targets) and by conventional tube and shell vial cell culture. One hundred fifty-eight confirmed (by cell culture and TK target PCR) positive and LightCycler-positive specimens were detected during the first 30 PCR cycles. LightCycler PCR-positive results for cycles 31 to 45 (39 of 67 [58.2%]) required confirmation by another PCR target (TK). LightCycler PCR is more sensitive (n = 197; 23.1%) than cell cultures (n = 150) for the routine laboratory detection of herpes simplex virus infections.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.