Abstract

Assessment of left ventricular ejection fraction (LVEF) is one of the primary objectives of echocardiography. The gold standard assessment technique in emergency medicine is eyeballing. A new tool is now available on pocket ultrasound devices (PUD): automatic LVEF. The primary aim of this study was to evaluate the concordance between LVEF values estimated by automatic LVEF with PUD and by cardiac magnetic resonance imaging (MRI). This was a prospective, monocentric, and observational study. All adult patients with an indication for cardiac MRI underwent a point-of-care ultrasound. Blinded to the MRI results, the emergency physician assessed LVEF using the automatic PUD tool and by visual evaluation. Sixty patients were included and analyzed. Visual estimation of LVEF was feasible for all patients and automatic evaluation for 52 (87%) patients. Lin's concordance correlation coefficient between automatic ejection fraction with PUD and by cardiac MRI was 0.23 (95% CI, 0.03-0.40). Concordance between LVEF estimated by the automatic ejection fraction with PUD and LVEF estimated by MRI was non-existent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.