Abstract
Objectives The vagus nerve stimulation device (VNS) is used for the management of seizures. This study evaluated what effect the diameter of the vagus nerve helical electrode might have on true vocal cord (TVC) mobility. The study was prompted after 2 cases of TVC immobility. Electrode nerve compression was suspect. Methods Eighteen patients underwent intraoperative vagus nerve measurement and electrode placement with subsequent voice and TVC evaluation. Electrode selection was based on vagus nerve measurements. Results Seven patients had vagus nerves measuring less than 2 mm diameter and received the 2-mm inner diameter electrode. Eleven patients had vagus nerves measuring more than 2 mm in diameter and received the 3-mm inner diameter electrode. No patients experienced transient or permanent hoarseness or paresis/paralysis. Conclusion Precise vagus nerve measurements and electrode selection appear to decrease the incidence of nerve compression injury and TVC immobility.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.