Abstract
This paper describes an inexpensive and relatively simple procedure to prepare high-performance size-exclusion columns from laboratory-packed or commercially pre-packed high-performance liquid chromatography columns by chemical deactivation of the surface of the silica packing material. The columns can be prepared quickly and reproducibly, they are chromatographically as efficient as high-performance size-exclusion columns commercially available, and are prepared at a fraction of the cost. Examples of the separation of proteins, peptides and polysaccharides on laboratory-prepared columns with differing permeation characteristics are illustrated and discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.