Abstract

Our aim was to evaluate the effects of intraaortic balloon pumping (IABP) on the blood velocity waveform in the absence or presence of coronary artery stenosis. Using anesthetized open-chest dogs, the septal arterial blood flow velocities were measured with a 20 MHz 80-channel ultrasound pulsed Doppler velocimeter in the absence (n = 5) or presence (n = 3) of left main coronary artery stenosis. The blood velocity waveform was analyzed by calculating the systolic retrograde velocity integral (SR) and the diastolic antegrade velocity integral (DA). A slosh ratio was defined as SR/DA. The left anterior descending arterial flow (CBF), aortic pressure (AoP), and poststenotic distal coronary pressure (DiP) were also measured simultaneously. We compared the effect of IABP on the velocity waveforms in the absence and in the presence of coronary artery stenosis. In the absence of stenosis, IABP increased DiP during diastole and augmented DA while it also increased SR. IABP augmented the net CBF because of the greater increase in DA than SR. In the presence of stenosis, however, IABP did not increase DiP and resulted in no significant effect on the net CBF.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.