Abstract

ObjectiveThis study aimed to examine the intra- and interlaboratory variations of cycle threshold (Ct) values using the nationwide proficiency testing for SARS-CoV-2.MethodsTriplicated strong-positive contrived samples duplicated weak-positive contrived samples, and 2 negative samples were transported to participating laboratories in October 2021.ResultsA total of 232 laboratories responded. All except 4 laboratories correctly answered. Six false-negative results, including 2 false-negatives with Ct values beyond the threshold and 1 clerical error, were noted from weak-positive samples. Intralaboratory variations of Ct values of weak-positive and strong-positive samples were not acceptable (Ct > 1.66) in 17 and 7 laboratories, respectively. High interlaboratory variations of Ct values (up to 7 cycles) for the 2 commonly used polymerase chain reaction (PCR) reagents were observed.ConclusionThe overall qualitative performance was acceptable; intralaboratory variation was acceptable. However, interlaboratory variations of Ct values were remarkable even when the same PCR reagents were used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.