Abstract

Because of the increased use of circular dichroism (CD) spectroscopy as a routine technique by nonspecialists to determine the conformational/configurational properties of biomolecules, we have decided to present here some criteria to accurately check the ordinate scale calibration of a CD spectrometer particularly in the critical low-wavelength UV region, to understand, and correct, where possible, the potential limitations coming from the hardware. We also analyze some wavelength calibration methods, and some standards for the CD-scale calibration, and we discuss the critical characteristics of current instrumentation affecting measurements. The example of the bovine catalase CD spectrum is considered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.