Abstract

We compared the performance of a new interferon gamma release assay (IGRA) format assay, the ichroma™ COVID-19 IGRA (IGRA-SARS), with that of thewidely used QuantiFERON SARS-CoV-2 ELISA kit (QFN-SARS) in vaccinated healthcare workers (HCWs). Additionally, we analyzed the long-term changes in IGRA results after the final vaccine dose. A total of 383 specimens from 281 HCWs were enrolled in this study, and the results of SARS-IGRA andQFN-SARS assays were compared. In addition, we performed the receive operator curve analysis to estimate the optimal cut-off value for IGRA-SARS. For all specimens, IGRA-SARS and QFN-SARS showed 75.7% and 64.2% of the positive results, respectively. The absolute agreement between IGRA-SARS and QFN-SARS was 80.0%, and the Fleiss' κ value was 0.525, indicating moderate agreement. ROC curve analysis of the IGRA-SARS results showed a cut-off value of >0.254IU/mL, which was consistent with the manufacturer's specifications. The positive rates of both IGRA assays decreased significantly after a postvaccination period of 6months. IGRA-SARS showed acceptable performance in the detection of vaccine-induced immunity against COVID-19; however, harmonization of IGRA assays has notyet been achieved. Additionally, the significant decline of positive rates of IGRA after the last vaccination wouldsupport the necessity of booster vaccination after a postvaccination period of 6months.