Abstract

Static gravimetric method with flying start-and-finish is widely used for calibration of large and medium liquid flow for its advantage of maintaining continuous flow through the flowmeter. However one needs to take careful measures in implementing this method in low flow systems to obtain good accuracy. At NMIJ/AIST, we adopted this calibration method using a pair of high speed switching valves as diverter for flow range below 1 L/h. This paper discusses the timing error of the diverting valves evaluated according to ISO4185 and also describes some of the main features of the gravimetric system using this high-speed switching valves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.