Abstract

The quantitative gated SPECT (QGS) software is able to calculate LV volumes and visualize LV wall motion and perfusion throughout the cardiac cycle using an automatic edge detection algorithm of the left ventricle. We evaluated the reliability of global and regional LV function assessment derived from QGS by comparing it with the results from left ventriculo-cineangiography (LVG). In 20 patients with left ventricular dysfunction who underwent ECG gated 99mTc-tetrofosmin SPECT, the end-diastolic volume (EDV), end-systolic volume (ESV) and ejection fraction (LVEF) were calculated. The QGS-assessed regional wall motion was determined using the cinematic display. QGS-derived EDV, ESV and LVEF correlated well with those by LVG (p<0.001 for each). There was a good correlation between wall motion score (WMS) derived from the QGS and the LVG (r=0.40, p<0.05). In some patients with extensive myocardial infarction, there was a discrepancy in the regional wall motion results between QGS and LVG. The ECG-gated SPECT using QGS is useful to evaluate global and regional LV functions in left ventricular dysfunction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.