Abstract

Background: GeneXpert has been used for Mycobacterium tuberculosis testing, but is currently available for HCV RNA testing. GeneXpert assay is expected to be more accurate, efficient, and cost-effective for HCV viral load quantification. This study intended to evaluate the new method quantification of plasma HCV RNA by the GeneXpert in comparison to the Roche Cobas TaqMan 96 as standard diagnostic tools.Method: A total of 54 HCV-infected plasma samples with anti-HCV positive were examined by GeneXpert assay, followed with Cobas TaqMan 96 for quantification of HCV RNA. Correlation was performed by Pearson test (in log10) and diagnostic test by Chi-square test. Sensitivity and specificity of the GeneXpert assay measured by calculating 2x2 contingency table. Bland-Altman plot were generated to assess the mean difference between the two assays.Results: GeneXpert has strong correlation to the Roche Cobas TaqMan 96 (R=0.993; P value 0.001). GeneXpert has sensitivity of 100% (95% CI: 90–100%) and specificity of 90% (95% CI: 54.1–99.5%). The Bland Altmand plot showed that one sampel has 1 log difference with the Roche Cobas TaqMan 96 measurement result.Conclusion: There was a strong correlation in the measurement of HCV RNA by GeneXpert and moreover its assay also has an excellent overall performance compared to Cobas TaqMan 96. Thus, it can be considered as a reliable tools for HCV virological response monitoring.

Highlights

  • Hepatitis C virus (HCV) infection is one of the leading causes of chronic liver disease in the world

  • The Bland Altmand plot showed that one sampel has 1 log difference with the Roche Cobas TaqMan 96 measurement result

  • There was a strong correlation in the measurement of Hepatitis c virus (HCV) RNA by GeneXpert and its assay has an excellent overall performance compared to Cobas TaqMan 96

Read more

Summary

Introduction

Hepatitis C virus (HCV) infection is one of the leading causes of chronic liver disease in the world. Besides of the baseline levels measurement, HCVinfected patients should undergo a treatment response monitoring by HCV RNAquantification. An accurate, efficient, and cost-effective diagnostic testing are needed to improve patients’ compliance in HCV infection diagnosis and therapy monitoring.[4,5] HCV RNA quantification is a confirmatory testing for anti-HCV antibodies result. Viral load is crucial parameter in detection and monitoring of HCV infection at all condition.[4,6,7]. GeneXpert assay is expected to be more accurate, efficient, and costeffective for HCV viral load quantification. This study intended to evaluate the new method quantification of plasma HCV RNA by the GeneXpert in comparison to the Roche Cobas TaqMan 96 as standard diagnostic tools

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.