Abstract
Hydroxyapatite (HA) coated metallic substrates are commonly used for load-bearing orthopedic implants. In the present work Titanium grade 2 and Titanium grade 5 substrates were coated with hydroxyapatite and hydroxyapatite reinforced with 5% (by wt.) multiwalled carbon nanotubes (MCNT) separately using plasma spray coating. The wear behavior of these coated substrates was studied with the simulated body fluid (SBF) having a pH between 7.20 to 7.40. Fracture toughness and wear resistance of HA reinforced with 5 % (by wt.) MCNT composite coating was found to be higher as compared to pure HA coating. This improvement is by virtue of inherent mechanical properties and the crack bridging effect offered by MCNT. The addition of MCNT also helped in restricting crack propagation in the coatings. As evident from SEM images, abrasive and adhesive wear mechanism were observed which reduces greatly with reinforcement of MCNT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.