Abstract
Near patient testing (NPT) and point-of-care testing (POCT) using portable benchtop analyzers has become necessary in many areas of the medical community, including biocontainment. We evaluated the Beckman AcT diff, Abaxis Vetscan HMII (two instruments), Abbott Cell-Dyn 1800, and Abaxis Vetscan VS2 for within-run precision and correlation to central laboratory instruments using non-human primates blood. Compared with the central laboratory instruments, the Beckman AcT diff correlated on 80%; the HMII instruments on 31% and 44%, the CD1800 on 31%, and the VS2 on 71% of assays. For assays with published manufacturers precision guidelines, the AcT diff met all nine, the HMII instruments met one and six of six, and the CD 1800 met one of six. Laboratories using NPT/POCT must test their individual instruments for precision and correlation, identify assays that are reliable, and exclude or develop supplemental procedures for assays that are not.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.