Abstract

Airway manikins have traditionally been used for teaching mask ventilation and tracheal intubation. There is an increasing need to use manikins for training in procedures such as insertion of the laryngeal mask airway. We have assessed four new airway training manikins (latest versions of the Airway Trainer (Laerdal, Norway), Airway Management Trainer (Ambu, UK), 'Bill 1' (VBM, Germany) and Airsim (Trucorp, Ireland)) as simulators for insertion of the LMA Classic laryngeal mask airway. Twenty volunteer anaesthetists inserted a size-4 laryngeal mask airway five times into each of the four manikins, in random order. Each insertion was assessed using objective and subjective tests. Subjective assessment varied widely but overall assessment indicated that the Airway Management Trainer was the poorest simulator for insertion of the laryngeal mask airway. The 'Bill 1' and Airsim manikins performed best as simulators for insertion of the laryngeal mask airway, although realistic ventilation with 'Bill 1' was not possible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.