Abstract

Abstract Vascular diseases like aneurysms or atherosclerosis belong to the leading causes of death. For diagnosis and treatment, medical imaging is essential. Standard imaging methods are Ultrasound, X-ray-based Angiography, Computed tomography or Magnet Resonant Imaging. Additionally, Intravascular Ultrasound or Optical Coherence Tomography are available for intravascular imaging. One imaging technique for direct investigation of the inner vessel is vascular endoscopy, also called angioscopy. It is based on the insertion of an optical endoscope into the blood vessel for direct imaging. Dependent on the vessel diameter, these endoscopes have to be very small. Angioscopy can provide valuable information about the condition of the vessel wall and serve for procedure observation or implant assessment. But angioscopy requires the absence of blood for a clear field of view. In prior research, we have already developed methods for the placement of a tiny optical fiber inside a catheter combination. As an essential step toward a clinical study, we now want to evaluate the flushing parameters needed to achieve a clear field of view for the endoscope. The parameters are affected by the blood flow and vessel diameter. We designed a test setup that allows angioscopic examination and the simulation of realistic conditions, such as blood flow and different vessel diameters and that includes a test pattern for image quality assessment.