Abstract

Bone densitometry using fan-beam technology allows both femurs to be rapidly measured with the aid of a bilateral foot-positioning tool to keep both legs in internal rotation. The Lunar Prodigy system uses a bilateral foot-positioning tool with a fixed interpedal width. The aim of this study was to determine whether the small degrees of femur abduction and adduction introduced by a fixed interpedal width-positioning tool influences the femur bone mineral density (BMD) measurements. Two-hundred patients had the total femur (FTot), femoral neck (FN), femoral shaft (FS), and femoral trochanter (FT) BMD, and angles of femoral abduction and adduction measured. Of all subjects, 26.5% had an absolute difference in femur angulation of 1 degrees , 77.5% had 5 degrees and 97% had 10 degrees. There was no significant correlation between the right-to-left differences in femur angles and the absolute right-to-left differences in FTot, FN, FS, and FT BMD. The introduction of small degrees of femur angulation with the fixed interpedal width bilateral foot positioner does not significantly affect the measured BMD at the various femur regions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.