Abstract

Background. Guidelines for the Diagnoses and Management of Asthma recommend that the peak flow meter should be used in a standing position. Objective. To determine whether the peak expiratory flow (PEF) varies with the position of the subject and to evaluate the impact of application of nose clip on PEF measurement. Methods. Using a Mini Wright Peak Flow Meter, the PEF was measured in 33 healthy adults in sitting and standing positions, with and without application of a nose clip. Each subject blew into the peak flow meter three times and best of the 3 blows was recorded. Results. The PEF was not significantly higher in standing than in sitting position. Application of nose clip had no significant impact on the PEF. Conclusion. The position of the subject and application of nose clip has no significant impact on PEF measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.