Abstract

The aim of this study was to evaluate entrance skin dose (ESD), organ dose and effective dose to patients undergoing catheter ablation for cardiac arrhythmias, based on the dosimetry in an anthropomorphic phantom. ESD values associated with mean fluoroscopy time and digital cine frames were in a range of 0.12-0.30 Gy in right anterior oblique (RAO) and 0.05-0.40 Gy in left anterior oblique (LAO) projection, the values which were less than a threshold dose of 2 Gy for the onset of skin injury. Organs that received high doses in ablation procedures were lung, followed by bone surface, esophagus, liver and red bone marrow. Doses for lung were 24.8-122.7 mGy, and effective doses were 7.9-34.8 mSv for mean fluoroscopy time of 23.4-92.3 min and digital cine frames of 263-511. Conversion coefficients of dose-area product (DAP) to ESD were 8.7 mGy/(Gy cm(2)) in RAO and 7.4 mGy/(Gy cm(2)) in LAO projection. The coefficients of DAP to the effective dose were 0.37 mSv/(Gy cm(2)) in RAO, and 0.41 mSv/(Gy cm(2)) in LAO projection. These coefficients enabled us to estimate patient exposure in real time by using monitored values of DAP.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.