Abstract
Abstract The aim of this study was to establish an indirect method to calculate the values of entrance surface air kerma in patients undergoing diagnostic examinations in X-ray systems with computed radiography based on the exposure index. The entrance surface air kerma values were compared with values obtained also indirectly based on measurements of X-ray tube output. The mean±standard deviation (1σ) and third quartile for entrance surface air kerma calculated from the exposure index were 2.1±1.0 mGy and 3.0 mGy, respectively. For entrance surface air kerma based on measurements of the X-ray tube output, the mean±standard deviation (1σ) and third quartile were respectively 3.1±1.9 mGy and 5.5 mGy. The observed values of entrance surface air kerma are smaller than the reference level adopted in Brazil (10 mGy). The results obtained with both methods were similar when taking into account the estimated uncertainties in the determination of air kerma values, although the reproducibility of the determinations based on the exposure index is better.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.