Abstract

A continuous-flow (timed injection) electrochemical hydride generation system was developed for the determination of total Sb in river water and sea-water. Lead was used as cathode material for the production of stibine. Both SbIII and Sbv were equally converted into their hydrides by electrochemical means with 92 ± 4% efficiency. The hydride was trapped in a Pd-coated graphite furnace prior to atomization. An absolute detection limit of 45 pg (3sblank) and a concentration detection limit of 0.02 µg l–1 were obtained using 2 ml sample volumes. The precision of replicate measurements at the 0.2 µg l–1 level (0.5 ng absolute mass level) was better than 6% relative standard deviation. The accuracy of the approach was verified through the analysis of National Research Council of Canada certified reference materials SLRS-2 and SLRS-3 River Water and NASS-4 Open Ocean Sea-water

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.