Abstract
Measurement of 222Rn is an important research concern because of the high-energy alpha emission of its 218Po and 214Po progeny. The LR-115 passive detector is the basis of a solid-state nuclear track detector (SSNTD) whose use for radon measurements has been well established. The usage of the SSNTD technique requires calibration by comparison. The aim of this work is to evaluate the efficiency calibration parameters for the radon measurement system that three steps: (1) irradiation conditions, (2) etching conditions, and (3) track counting. The following parameters were evaluated: breakdown and counting voltages for the spark counter, detector thickness, etching time, temperature and NaOH concentration as etching conditions for the LR-115 detector. Calibrating factors of 0.035 and 0.028 tracks cm-2 Bq-1 m3 d-1 for the open- and closed-chamber radon monitors, respectively, were determined considering the best efficiency calibration parameter values.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.