Abstract

BackgroundData from literature shows that antigen tests are rapid and helpful tools for diagnosis of COVID-19.AimThis work aimed to evaluate the performances of the Elecsys SARS-CoV-2 Antigen test, in comparison to RT-qPCR, the gold standard.MethodsA total of 110 swabs were tested; according to rRT-PCR, 76 were positive, and 34 were negative. The swabs were processed by Elecsys SARS CoV 2 Antigen assay (Roche Diagnostics GmbH, Mannheim, Germany), an electrochemiluminescence immunoassay (ECLIA).ResultsIn a first evaluation, the overall sensitivity and specificity were 85% and 100%, respectively. It was noted that most of the discordant cases had cycle threshold (Ct) values > 28. Therefore, it was assumed a new measure to evaluate sensitivity and specificity, then samples with Ct values < 28 were selected. In this way, it was achieved a Ct < 28 sensitivity of 94%. The level of agreement between the two tests was 89. 1% with κ value of 0.77 for total data and 95.9% with κ value of 0.95 for samples with < 28 Ct. The antigen test performs well in the presence of high viral loads, whereas lower levels are missed.ConclusionsThe comparison data obtained in this study support that this method seems a proper approach for rapid screening of patients with high SARS-CoV-2 viral load; however, the rate of sensitivity is highly Ct-dependent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.