Abstract

Detailed measures of infant body composition are needed for understanding the impact of genes and environment on growth early in life. The purpose of this study was to compare the accuracy and bias of body composition in infants. Dual energy X-ray absorptiometry (DXA) and magnetic resonance imaging (MRI) were used to determine body composition and the trunk depot. The depots measured were total fat mass (FM), total fat-free mass (FFM) and trunk FM and FFM using DXA and MRI in 14 infants. None of the regression lines between DXA and MRI significantly deviate from the line of identity for any of the depots studied. However, Bland-Altman analyses revealed bias for trunk FM and trunk FFM. Our data showed DXA to be accurate (regression not significantly deviating from the line of identity), with high agreement (indicated by high R(2) ) and without bias (non-significant Bland-Altman) when estimating total FM and FFM. This could not be said for trunk estimates.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.