Abstract

Butylatedhydroxytoluene (BHT) rods represent a potential dosimeter in radiation processing, with readout via electron paramagnetic resonance (EPR) spectroscopy. Among the possible sources of uncertainty are those associated with the performance of the dosimetric medium and the conditions under which measurements are made, including sampling and environmental conditions. Present study makes estimate of the uncertainties, investigating physical response in different resonance regions. BHT, a white crystalline solid with a melting point of between 70–73°C, was investigated using 60Co gamma irradiation over the dose range 0.1–100kGy. The intensity of the EPR signal increases linearly in the range 0.1–35kGy, the uncertainty budget for high doses being 3.3% at the 2σ confidence level. The rod form represents an excellent alternative dosimeter for high level dosimetry, of small uncertainty compared to powder form.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.