Abstract

We demonstrate agreement between measurements and ab initio calculations of the frequency shifts caused by distributed cavity phase variations in the microwave cavity of a primary atomic fountain clock. Experimental verification of the finite element models of the cavities gives the first quantitative evaluation of this leading uncertainty and allows it to be reduced to δν/ν=±8.4×10(-17). Applying these experimental techniques to clocks with improved microwave cavities will yield negligible distributed cavity phase uncertainties, less than ±1×10(-17).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.