Abstract

DNA was extracted from eleven bone samples using PrepFiler Express BTA™ Forensic DNA Extraction kit with AutoMate Express™ Forensic DNA Extraction System with 20 and 50 μL of elution volume. The objective was to compare extraction efficiency with different elution volumes, evaluating DNA recovery, profile quality and observing if sample concentration alters the degradation index (DI) and the presence of inhibitors. The analyzed samples showed increased DNA levels with the elution volume of 20 μL, and the inhibition values did not change among the protocols, however some changes in the DI were observed. Undesirably, for low DNA concentrations, the 20 μL protocol demands more extractions per sample. This 20 μL method proved to be useful for difficult samples with low levels of DNA and high degradation, being able to increase DNA concentration and obtain genetic profiles with higher quality compared to the protocol in use, increasing the efficiency of analysis, important in cases of disaster victim identification (DVI).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.