Abstract

Flow volumes in an in vitro model of the aorta with 3 different degrees of stiffness (stiff, moderately stiff, and compliant) proximal to a coarctation were calculated by using a digital color Doppler echocardiography flow calculation method that semiautomatically integrates spatial and temporal color flow velocity data. These flow volumes were compared with those obtained by the conventional pulsed Doppler method with reference to ultrasonic flowmeter. Flow volumes determined by the automated method agreed well with those obtained by ultrasonic flowmeter, even in this compliant aorta model with vessel size changing with pulsation, whereas the pulsed Doppler method overestimated the reference data, especially for more compliant descending aortic segments. The combination of flow data with continuous wave Doppler allows definition of effective orifice area for coarctation. (J Am Soc Echocardiogr 1999;12:517-26.)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.