Abstract
This study assessed the influence of deep learning reconstruction (DLR) on the quality of diffusion-weighted images (DWI) and apparent diffusion coefficient (ADC) using an ice-water phantom. An ice-water phantom with known diffusion properties (true ADC = 1.1 × 10-3 mm2/s at 0°C) was imaged at various b-values (0, 1000, 2000, and 4000s/mm2) using a 3T magnetic resonance imaging scanner with slice thicknesses of 1.5 and 3.0mm. All DWIs were reconstructed with or without DLR. ADC maps were generated using combinations of b-values 0 and 1000, 0 and 2000, and 0 and 4000s/mm2. Based on the quantitative imaging biomarker alliance profile, the signal-to-noise ratio (SNRs) in DWIs was calculated, and the accuracy, precision, and within-subject parameter variance (wCV) of the ADCs were evaluated. DLR improved the SNR in DWIs with b-values ranging from 0 to 2000s/mm2; however, its effectiveness was diminished at 4000s/mm2. There was no noticeable difference in the ADCs of images generated with or without implementing DLR. For a slice thickness of 1.5mm and combined b-values of 0 and 4000s/mm2, the ADC values were 0.97 × 10-3and 0.98 × 10-3mm2/s with and without DLR, respectively, both being lower than the true ADC value. Furthermore, DLR enhanced the precision and wCV of the ADC measurements. DLR can enhance the SNR, repeatability, and precision of ADC measurements; however, it does not improve their accuracies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.